| Both sides previous revisionPrevious revisionNext revision | Previous revision |
| observation:2027campaign [2026/11/23 15:30] – etienne | observation:2027campaign [2026/11/23 15:35] (current) – [Available instrumentation for 2027] etienne |
|---|
| === MTR2 in spectroscopy mode === | === MTR2 in spectroscopy mode === |
| In this mode, up to 4 (new) Andor Zyla cameras up to 4 (old) EMCCD Andor iXon camera. Because of informatics and optical elements constraints, a maximum of 5 cameras can be setup and used simultaneously (with their respective specification), in accordance with the MTR2 spectrograph output physical limitations. | In this mode, up to 4 (new) Andor Zyla cameras up to 4 (old) EMCCD Andor iXon camera. Because of informatics and optical elements constraints, a maximum of 5 cameras can be setup and used simultaneously (with their respective specification), in accordance with the MTR2 spectrograph output physical limitations. |
| * The Zyla cameras have 2048x2048 pixels: a typical setup shall give a spatial pixel of ~0.06" and a corresponding spectral pixel about ~3mÅ/pixel, which is much lower than the spectral resolution of MTR2 (hence we suggest to rebin, by a factor of 2 at least, in the lambda direction to increase S/N ratio). | * The Zyla cameras have 2048x2048 pixels: a typical setup shall give a spatial pixel of ~0.06” and a corresponding spectral pixel about ~3mÅ/pixel, which is much lower than the spectral resolution of MTR2 (hence we suggest to rebin, by a factor of 2 at least, in the lambda direction to increase S/N ratio). |
| * The Ixon cameras have 512x512 pixels : a typical setup on these camera with a standard de-magnification will give a spectral resolution of ~0.0123 Å/pixel (12.3 mÅ/pixel) which is roughly the MTR2 spectral resolution. The spatial pixel will be about ~0.25”/pixel. | * The Ixon cameras have 512x512 pixels : a typical setup on these camera with a standard de-magnification will give a spectral resolution of ~0.0123 Å/pixel (12.3 mÅ/pixel) which is roughly the MTR2 spectral resolution. The spatial pixel will be about ~0.25”/pixel. |
| |
| === MTR2 in spectropolarimetry mode: === | === MTR2 in spectropolarimetry mode: === |
| THEMIS polarimetric analysis scheme is based on a full-Stokes analyzer located at the F1 prime focus. For any solar polarization state to measure, the output of the analyzer is made of dual superimposed complementary linear polarizations. Thanks to THEMIS "polarization friendly" new optical scheme, this output travels through the telescope and arrives unperturbed on the spectrograph cameras [[technical:mtr2 |(cf. information about MTR2 spectrograph)]]. Just in front of the camera, a Wollaston splitter separates the beam into complementary Stokes components to form the spectral focal plane. THEMIS own three Wollaston, hence 3 cameras/3 wavebands at most can be simultaneously observed with polarimetric signal, among the 6 spectral cameras that THEMIS proposes. It is however possible to combine polarimetric and non-polarimetric cameras. | THEMIS polarimetric analysis scheme is based on a full-Stokes analyzer located at the F1 prime focus. For any solar polarization state to measure, the output of the analyzer is made of dual superimposed complementary linear polarizations. Thanks to THEMIS "polarization friendly" new optical scheme, this output travels through the telescope and arrives unperturbed on the spectrograph cameras [[technical:mtr2 |(cf. information about MTR2 spectrograph)]]. Just in front of the camera, a Wollaston splitter separates the beam into complementary Stokes components to form the spectral focal plane. THEMIS own three Wollaston, hence 3 cameras/3 wavebands at most can be simultaneously observed with polarimetric signal, among the 6 spectral cameras that THEMIS proposes. It is however possible to combine polarimetric and non-polarimetric cameras. |
| | |
| Regarding the available analysis software: | |
| | |
| === Broadband imaging (BBI) context camera: === | === Broadband imaging (BBI) context camera: === |
| === Available analysis softwares === | === Available analysis softwares === |
| |
| * Knox-Thompson image reconstruction for BBI | * Broadband imaging (BBI): Knox-Thompson (speckle) image reconstruction on-site tools. |
| * Reconstructed spectral map quick-looks for on site analysis, to be adapted to specific wavelength. | * MTR2: a python MTR2 data reduction and data inspection software package, either with or without polarimetry: [[https://github.com/SaidaKIS/Themis.git]] |
| * Stokes map quick-looks | * MTR2 reconstructed spectral map quick-looks for on-site analysis, to be adapted to specific wavelength. |
| * Thanks to V. Bommier, a basic magnetic field inversion software package (running locally in IDL) is also available. | * MTR2 Stokes map quick-looks for on-site analysis. |
| | * If observing the Fe I @ 6302 \AA spectral range with MTR2, thanks to V. Bommier, a basic magnetic field inversion software package (Matlab & Fortran), based on the Unnofit code, is also available, as well as quicklook magnetograms. |
| | * IBIS 2.0: xxxxx |
| |
| |